ECG Amplifier By TLC274
To make this adjustment, connect both inputs of the instrumentation amplifier together, and then connect a 100mV, 50Hz AC signal between the connected inputs and earth. Measure the output signal using an oscilloscope, and adjust P1 to minimize the level of the output signal. It is important that the electrodes make very good contact with the skin. In our test measurements, winding three uninsulated copper wires several times around the index fingers (and the right leg) proved to be sufficient to provide a good signal. The amplitude of the ECG signal measured with this arrangement was 200mV. The current consumption of this circuit is only 2mA, so the batteries should last a long time. This circuit must never be connected to a mains-operated power supply, in consideration of safety precautions that are necessary when making this sort of measurement on the human body.